top of page

The LR-Cal FLUID 100-45 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

  • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
  • Ramp function
  • Thermostat tests
  • Computer-controlled calibrations via computer interface
  • Homogeneous temperature distribution due to adjustable magnetic mixer
  • Option: dry block inserts

 

Version LR-Cal FLUID 100-45

  • Standard version
  • Order-Code FLUID100-45
  • Without inputs for resistance thermometer / thermocouples

Version LR-Cal FLUID 100-45-2I

  • Version with 2 measuring inputs for resistance thermometers Pt 100 (2-, 3- or 4-wire) or Pt 1000 as well as thermocouples incl. cold junction compensation types B, E, J, K, N, R, S and T.
  • Order-Code FLUID100-45-2I
  • The signals from up to two external temperature sensors can also be displayed.
  • Extended scope of delivery:
  • Electrical connection cables, plugs/terminals for the electrical connections.

 

Accuracy (maximum deviation) of the two measuring inputs for device version "-2I":

 

Pt 100 or Pt 1000at -40°C:±0.09°Cat 0°C:±0.08°Cat +150°C:±0.11°Cat +300°C:±0.14°C
Thermocouple type Bat +950°C:±0.97°Cat +1050°C:±1.03°Cat +1200°C:±1.12°C  
Thermocouple type Eat -40°C:±0.42°Cat 0°C:±0.40°Cat +350°C:±0.61°C  
Thermocouple type Jat +200°C:±0.52°Cat +450°C:±0.67°Cat +700°C:±0.82°C  
Thermocouples types K, N, R and Sat +400°C:±0.64°Cat +700°C:±0.82°Cat +1000°C:±1.00°C  
Thermocouple type Tat -40°C:±0.42°Cat 0°C:±0.40°Cat +350°C:±0.61°C 

LR-Cal FLUID 100-45/100-45-2I Portable Temperature Calibration Bath -35...+140°C

$10,449.00Price
Quantity
  • Description: Data:
    Temperature range (at 20°C ambient temperature) -35°C...+140°C
    Temperature indication Accuracy±0.15°C
    Resolution 0.01°/0.1° (°C or °F)
    Stability of regulated temperature ±0.02°C
    Heating from ambient temperature to 140°C, incl. stabilization approx. 40 minutes
    Cooling from 140°C to ambient temperature, incl. stabilization approx. 100 minutes
    Uniformity of temperature at 40 mm insertion depth radial: ±0.02°C at 0°C; axial: ±0.05°C
    Calibration fluid Silikone oil 200C5
    Reservoir Depth 170 mm x diameter 45 mm
    Power supply 230 VAC
    Power consumption 350 VA
    Interface RS232 (USB converter available as option)
    Fluid reservoir material Aluminium
    Fluid reservoir volume approx. 500 cm³
    Housing material metal
    Weight approx. 11 kg
    Dimension 160 x 360 x 350 mm
    Order-Code (standard version) FLUID100-45
    Order-Code (-2I version with 2 measuring inputs) FLUID100-45-2I