The LR-Cal FLUID 100-45 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.
- Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
- Ramp function
- Thermostat tests
- Computer-controlled calibrations via computer interface
- Homogeneous temperature distribution due to adjustable magnetic mixer
- Option: dry block inserts
Version LR-Cal FLUID 100-45
- Standard version
- Order-Code FLUID100-45
- Without inputs for resistance thermometer / thermocouples
Version LR-Cal FLUID 100-45-2I
- Version with 2 measuring inputs for resistance thermometers Pt 100 (2-, 3- or 4-wire) or Pt 1000 as well as thermocouples incl. cold junction compensation types B, E, J, K, N, R, S and T.
- Order-Code FLUID100-45-2I
- The signals from up to two external temperature sensors can also be displayed.
- Extended scope of delivery:
- Electrical connection cables, plugs/terminals for the electrical connections.
Accuracy (maximum deviation) of the two measuring inputs for device version "-2I":
Pt 100 or Pt 1000 | at -40°C: | ±0.09°C | at 0°C: | ±0.08°C | at +150°C: | ±0.11°C | at +300°C: | ±0.14°C |
Thermocouple type B | at +950°C: | ±0.97°C | at +1050°C: | ±1.03°C | at +1200°C: | ±1.12°C | ||
Thermocouple type E | at -40°C: | ±0.42°C | at 0°C: | ±0.40°C | at +350°C: | ±0.61°C | ||
Thermocouple type J | at +200°C: | ±0.52°C | at +450°C: | ±0.67°C | at +700°C: | ±0.82°C | ||
Thermocouples types K, N, R and S | at +400°C: | ±0.64°C | at +700°C: | ±0.82°C | at +1000°C: | ±1.00°C | ||
Thermocouple type T | at -40°C: | ±0.42°C | at 0°C: | ±0.40°C | at +350°C: | ±0.61°C |
LR-Cal FLUID 100-45/100-45-2I Portable Temperature Calibration Bath -35...+140°C
$10,449.00Price
Description: Data: Temperature range (at 20°C ambient temperature) -35°C...+140°C Temperature indication Accuracy±0.15°C
Resolution 0.01°/0.1° (°C or °F)Stability of regulated temperature ±0.02°C Heating from ambient temperature to 140°C, incl. stabilization approx. 40 minutes Cooling from 140°C to ambient temperature, incl. stabilization approx. 100 minutes Uniformity of temperature at 40 mm insertion depth radial: ±0.02°C at 0°C; axial: ±0.05°C Calibration fluid Silikone oil 200C5 Reservoir Depth 170 mm x diameter 45 mm Power supply 230 VAC Power consumption 350 VA Interface RS232 (USB converter available as option) Fluid reservoir material Aluminium Fluid reservoir volume approx. 500 cm³ Housing material metal Weight approx. 11 kg Dimension 160 x 360 x 350 mm Order-Code (standard version) FLUID100-45 Order-Code (-2I version with 2 measuring inputs) FLUID100-45-2I