top of page

The MO-5898-100A high current micro-ohmmeter employs the 4 terminals-method to avoid measurement errors caused by test leads and their contact resistances. Resistances readings are shown in the alphanumeric display with up to 4½ digits-resolution. It allows to measure resistances up to 200 Ω, with a resolution of up to 0.1 µΩ.

 

Using its internal rechargeable battery, measurements with up to 10 A test current may be carried out without connecting the equipment to mains power. Powered by to mains power, it allows to measure with any test current up to 100 A.

 

Measurement accuracy is guaranteed by an state-of-the-arts signals amplification system, offset-free and of high long-term stability.

 

Test current may be adjusted by the operator in every one of the scales and their values are measured using an analogue indicator (bargraph), making it easy to measure resistances with a significant inductive component, as in the case of big transformers windings.

 

The high-current generation system is based on modern technology that allows to significantly decrease both its weight (14 kg) and size. The cabinet is made of plastic material highly resistant to impacts and to environmental challenges.

Internal thermal sensors in all sensitive components avoid any damaged caused to the instrument due to overheating.

This is a strong but lightweight equipment, and may be easily carried by one person. It is water-resistant and can be used under severe weather conditions offering an excellent performance working both in the laboratory and out in the field.

Tinsley MO-5898-100A 100 Amp Digital Micro-Ohmmeter (1mA to 100A) 200 Ω DLRO

US$5,549.00Price
Quantity
  • Features

    • Alphanumerical display
    • Resolution: 0.1 μΩ
    • Resistance reading: up to 200 Ω
    • Up to 100 A test current
    • Kelvin-type (4-wires) measurement
    • Powered by internal battery (up to 10 A) or mains supply
    • Direct reading (Up to 4½ digits)
    • Overheating Protection
    • USB interface
  • Specifications

    TEST CURRENTS
    1 mA, 10 mA, 100 mA, 1 A, 10 A, 100 A.
    Each current may be continuously adjustable from 0 to
    100%.
    RESISTANCE RANGES
    Current Range Resolution
    100 A 0-199.9 μΩ 0.1 μΩ
    100 A 0-1999 μΩ 1 μΩ
    10 A 0-1999 μΩ 1 μΩ
    10 A 0-19.999 mΩ 1 μΩ
    1 A 0-19.99 mΩ 10 μΩ
    1 A 0-199.99 mΩ 10 μΩ
    100 mA 0-199.9 mΩ 100 μΩ
    100 mA 0-1999.9 mΩ 100 μΩ
    10 mA 0-1999 mΩ 1 mΩ
    10 mA 0-19.999 Ω 1 mΩ
    1 mA 0-19.99 Ω 10 mΩ
    1 mA 0-199.99 Ω 10 mΩ
    For each test current, ranges are automatically
    selected for optimal reading.
    MEASUREMENT PRINCIPLE
    Four-terminal, Kelvin-type.
    CONTINUOUS OPERATION TIME
    At 100 A this equipment may be used continuously for
    aprox. 15 minutes before the thermal protection
    activates.
    At 10 A or less, there is not a limited time for
    continuous operation.
    BASIC ACCURACY
    R < 0.5 mΩ: ± (0.50% of reading + 2 ULSD*).
    R ≥ 0.5 mΩ: ± (0.20% of reading + 2 ULSD*).
    * Units of the Least Significant Digit.

  • Order Info

    Please allow 4 - 5 weeks lead time for this new product to arrive. 

    Contact us for quote and more info about this product or other Tinsley offerings.

  • Included Accessories

    • Standard Scope of Delivery Includes:
    • 2 Combined current and potential leads for 10 A.
    • 2 Combined current and potential leads for 100 A.
    • 1 Power cord.
    • 1 USB cable.
    • 1 User guide.
    • 1 Case for the accessories.

Related Products

bottom of page