The LR-Cal LTC-DB-0600-35 metal dry block temperature calibrator serves as both a temperature source and reference device for testing and calibrating all types of temperature measuring instruments.
- Temperature range Ambient temperature upto+600°C
- Homogeneous temperature distribution
- Easy operation with color touchscreen
- Ramp function + thermostat tests
- Computer-controlled calibrations via USB interface and optional LR-Cal Aq2Sp2 software for MS Windows® PCs
- Numerous block inserts with different bores available for different test objects
- Option: Block insert with “black body” for calibrating infrared thermometers
Less thermal stress for your test objects:
The innovative ventilation concept of the LR-Cal LTC-DB-0600-35 significantly reduces the temperature on the top of the calibrator. A targeted tangential air flow passes evenly through the device and directs the warm air to the rear in a controlled manner. This keeps the measuring heads and housings of your test objects in a noticeably cooler environment. Measurement and compensation errors that can arise due to excessive ambient temperatures are thus effectively minimized.
LR-Cal LTC-DB-0600-35 Metal Dry Block Temperature Calibrator Ambient Temp +600°C
Features
- Portable metal dry block temperature calibrator
- Range-50 to +120°C
- Accuracy ±0.15°C, stability ±0.03°C
- Dry block: 35 mm diameter
Immersion depth: 130 mm
Included Accessories
- Portable metal block temperature calibrator LR-Cal LTC-DB-0600-35
- Spare fuses
- Power supply cable
- Connection cable for thermostat tests
- 1 block insert with 5 holes: 3.5 + 5.0 + 6.5 + 8.5 + 12.5 mm diameter
- Only if option LTC-MP-3I is ordered: Electrical connection cables and connectors red/black for the three measurement inputs
- Acceptance test certificate 3.1 according to DIN EN 10204 (factory calibration certificate)
- Operating instructions DE+EN
Optional accessories
- 3 configurable measuring inputs
- Power supply 115 VAC
- DAkkS calibration certificate
- PC software Aq2Sp2
- Carrying bag with shoulder trap
- Sturdy aluminum case
- Removal tool for block insert
Ordering information
This item comes with a one year warranty from the manufacturer.
Please allow 5 - 6 weeks lead time for this new product to arrive.
Specifications
Description: Data: Temperature range: ambient temperature upto+600°C Display: Touchscreen (color) Display accuracy: ±0.3°C full temperature range * Units of measure: °C / °F / K Display resolution: 0,1° / 0,01° / 0,00^° Mean heating time:
(incl. stabilization)approx. 40 min.
from ambient temperature to +550°CMean cooling time:
(incl. stabilization)approx. 60 min.
from +550°C to +100°CStability: ±0.05°C full temperature range ** Axial uniformity: ±0.15°C at +400°C *** Radial uniformity: ±0.35°C at +400°C Hole (block) diameter: 35 mm Hole (block) depth: 185 mm Insert material: nickel plated brass Switch test; voltage: Open/Closed; 12 VDC Adjustable ramp function: from 0.1°C/min PC interface: USB type B Automatic calibration: at 5 temperature points External probes: Optional: Order-Code LTC-MP-3I (to be ordered together with device) Operating voltage: 230 VAC (optional 115 VAC) 50/60 Hz Electric power: 800 W Dimensions: 160 x 360 x 350 mm Weight incl. std. accessories: 10.0 kg All values have been measured at ambient temperature of +20°C.
The achievement of stabilization is confirmed by a displayed root symbol and an audial beep.
* = Temperature deviation between display and reference probe.
** = Max. temperature difference at stable temperature over 30 min.
*** = Measured from the button up to 40 mm.Optional Inserts
Outer dimensions: 35 x 190 mm, usable depth 185 mm
Order-Code Description LTC-BI-35-190-0 Dry block insert without holes (for self-drillings) LTC-BI-35-190-5A Dry block insert with 5 holes: 3.5 + 5.0 + 6.5 + 8.5 + 12.5 mm diameter
1 pc. already included in the standard scope of delivery of this calibratorLTC-BI-35-190-5B Dry block insert with 5 holes: 6.5 + 6.5 + 6.5 + 6.5 + 6.5 mm diameter LTC-BI-35-190-5C Dry block insert with 5 holes: 5.0 + 5.0 + 5.0 + 5.0 + 6.5 mm diameter LTC-BI-35-190-C2 Dry block insert with 2 holes as per customers request (each 3.5...20 mm diameter) LTC-BI-35-190-C3 Dry block insert with 3 holes as per customers request (each 3.5...20 mm diameter) LTC-BI-35-190-C4 Dry block insert with 4 holes as per customers request (each 3.5...20 mm diameter) LTC-BI-35-190-C5 Dry block insert with 5 holes as per customers request (each 3.5...20 mm diameter) LTC-BI-35-190-C6 Dry block insert with 6 holes as per customers request (each 3.5...20 mm diameter) LTC-BI-35-190-C9 Dry block insert with 9 holes as per customers request (each 3.5...7 mm diameter) LTC-BI-35-190-BB Dry block insert with "Black Body", conical, 26 to 15 mm inner diameter, for infrared thermometers
NOTE:
For calibrator without option LTC-MP-3I a reference thermometer LR-Cal LRT 750 is needed.
For devices with option LTC-MP-3I, a Pt 100 reference sensor class A, order-code BB-SENSOR-01 is required.