The LR-Cal LTC-FB-9035-45 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.
- Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
- Homogeneous temperature distribution due to adjustable magnetic mixer
- Simple operation with color touchscreen
- Ramp function + thermostat tests
- Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
- Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
- Option: dry block inserts to convert the device to a dry block temperature calibrator
LR-Cal LTC-FB-9035-45 Temperature Calibration Micro Bath -35...+120°C
Description: Specification: Temperature range: -35...+120°C Display + operation: color touchscreen Display accuracy: ±0.15°C * Measuring units: °C / °F / K Displey resolution: 0.1° / 0.01° / 0.00^° Mean heating time:
(incl. stabilization)approx. 40 min.
from ambient temperature to +120°CMean cooling time:
(incl. stabilization)approx. 100 min.
from +120°C to ambient temperatureStability: ±0.02°C at 0°C ** Axial uniformity: ±0.05°C Radial uniformity: ±0.02°C at 0°C *** Hole diameter: 45 mm Hole depth: 170 mm Insert material: anticorodal Switch test; voltage: Open/Closed; 4.5 VDC Adjustable ramp function: from 0.1°C/min Computer interface: USB type B Automatic calibration: at 5 temperature points External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument) Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz Electric powert: 350 VA Dimensions: 160 x 360 x 350 mm Weight incl. standard accessories: 11.0 kg All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
* = Temperature deviation between display and reference probe.
** = Max. temperature difference at stable temperature over 30 min.
*** = Measured from the button up to 40 mm.