top of page

The LR-Cal LTC-DB-9030-35 metal dry block temperature calibrator serves as both a temperature source and reference device for testing and calibrating all types of temperature measuring instruments.

  • Portable metal dry block temperature calibrator
  • Range -30 to +150°C
  • Accuracy ±0.15°C, stability ±0.03°C
  • Dry block: 35 mm diameter
    Immersion depth: 130 mm

 

Option LTC-MP-3I - three configurable measuring inputs

The LTC-MP-3I option can be ordered together with the LR-Cal LTC-DB-9030-35:

 

Device version with 3 configurable measuring inputs. Of these, 2 for resistance thermometers (Pt 100 (2-, 3- or 4-wire) or Pt 100, or thermocouples incl. cold junction compensation, types  B, E, J, K, N, R, S and T, and 1 measuring input for analog output signals 4 to 20 mA or 0 to 10 V.
Order-Code: LTC-MP-3I (must be ordered together with the base unit)

 

Extended scope of delivery:
Electrical connection cables, plugs/terminals (red/black) for electrical connections

Accuracy (max. deviation) of resistance and thermocouple measurement inputs with option LTC-MP-3I:
Pt 100 or Pt 1000at -40°C:±0.09°Cat 0°C:±0.08°Cat +150°C:±0.11°Cat +300°C:±0.14°C
Thermocouple type Bat +950°C:±0.97°Cat +1050°C:±1.03°Cat +1200°C:±1.12°C  
Thermocouple type Eat -40°C:±0.42°Cat 0°C:±0.40°Cat +350°C:±0.61°C  
Thermocouple type Jat +200°C:±0.52°Cat +450°C:±0.67°Cat +700°C:±0.82°C  
Thermocouple types K, N, R and Sat +400°C:±0.64°Cat +700°C:±0.82°Cat +1000°C:±1.00°C  
Thermocouple type Tat -40°C:±0.42°Cat 0°C:±0.40°Cat +350°C:±0.61°C 

 

 

LR-Cal LTC-DB-9030-35 Metal Dry Block Temperature Calibrator -30 to+150°C

US$7,309.00Price
Quantity
  • Features

    • Portable metal dry block temperature calibrator
    • Range-50 to +120°C
    • Accuracy ±0.15°C, stability ±0.03°C
    • Dry block: 35 mm diameter
      Immersion depth: 130 mm
  • Included Accessories

    • Portable metal block temperature calibrator LR-Cal LTC-DB-9030-35
    • Spare fuses
    • Power supply cable
    • Connection cable for thermostat tests
    • 1 block insert with 6 holes: 3.5 + 4.5 + 5.5 + 6.5 + 8.5 + 10.5 mm diameter
    • Only if option LTC-MP-3I is ordered: Electrical connection cables and connectors red/black for the three measuring inputs
    • Acceptance test certificate 3.1 according to DIN EN 10204 (factory calibration certificate)
    • Operating instructions DE+EN
  • Optional accessories

    • 3 configurable measuring inputs
    • Power supply 115 VAC
    • DAkkS calibration certificate
    • PC software Aq2Sp2
    • Carrying bag with shoulder trap
    • Sturdy aluminum case
    • Removal tool for block insert
  • Ordering information

    This item comes with a one year warranty from the manufacturer.

    Please allow 5 - 6 weeks lead time for this new product to arrive. 

  • Specifications

    Description: Data:
    Temperature range: -30 to +150°C
    Display: Touchscreen (color)
    Display accuracy: ±0.15°C full temperature range *
    Units of measure: °C / °F / K
    Display resolution: 0,1° / 0,01° / 0,00^°
    Mean heating time:
    (incl. stabilization)
    approx. 28 min.
    from +20°C to +140°C
    Mean cooling time:
    (incl. stabilization)
    approx. 23 min.
    from +20°C to -25°C
    Stability: ±0.03°C **
    Axial uniformity: ±0.01°C
       
    Hole (block) diameter: 35 mm
    Hole (block) depth: 135 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4,5 VDC
    Adjustable ramp function: from 0.1°C/min
    PC interface: USB type B
    Automatic calibration: at 5 temperature points
    External probes: Optional: Order-Code LTC-MP-3I (to be ordered together with device)
    Operating voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric power: 350 W
    Dimensions: 160 x 350 x 360 mm
    Weight incl. std. accessories: 9.3 kg

    All values have been measured at ambient temperature of +20°C.
    The achievement of stabilization is confirmed by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 60 mm.

  • Optional Inserts

    Order-Code Description 
    LTC-BI-35-135-0 Dry block inserts without holes (for self-drillings)
    LTC-BI-35-135-6 Dry block insert with 6 holes: 3.5 + 4.5 + 5.5 + 6.5 + 8.5 + 10.5 mm diameters
    1 pc. already included in standard scope of delivery of the calibrator
    LTC-BI-35-135-2 Dry block inserts with 2 holes: 6.5 + 19.5 mm diameters
    LTC-BI-35-135-C1 Dry block insert with 1 to 3 holes as per customers request (each 3.5...20 mm diameter)
    LTC-BI-35-135-C4 Dry block insert with 4 to 6 holes as per customers request (each 3.5...20 mm diameter)
    LTC-BI-35-135-C7 Dry block insert with 7 to 10 holes as per customers request (each 3.5...20 mm diameter)
    LTC-BI-35-135-BB Dry block insert with "Black Body", 26 mm inner diameter, for infrared thermometer
    NOTE:
    For calibrator without option LTC-MP-3I a reference thermometer LR-Cal LRT 750 is needed.
    For devices with option LTC-MP-3I, a Pt 100 reference sensor class A, order-code BB-SENSOR-01 is required.