top of page

The LR-Cal LTC-FB-9018-60 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

  • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
  • Homogeneous temperature distribution due to adjustable magnetic mixer
  • Simple operation with color touchscreen
  • Ramp function + thermostat tests
  • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
  • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
  • Option: dry block inserts to convert the device to a dry block temperature calibrator

LR-Cal LTC-FB-9018-60 Temperature Calibration Micro Bath -18...+140°C

Quantity
  • Description: Specification:
    Temperature range: -18...+140°C
    Display + operation: color touchscreen
    Display accuracy: ±0.15°C *
    Measuring units: °C / °F / K
    Displey resolution: 0.1° / 0.01° / 0.00^°
    Mean heating time:
    (incl. stabilization)
    approx. 50 min.
    from ambient temperature to +140°C
    Mean cooling time:
    (incl. stabilization)
    approx. 50 min.
    from +140°C to ambient temperature
    Stability: ±0.02°C at 0°C **
    Axial uniformity: ±0.05°C
    Radial uniformity: ±0.07°C at +140°C ***
    Hole diameter: 60 mm
    Hole depth: 170 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4.5 VDC
    Adjustable ramp function: from 0.1°C/min
    Computer interface: USB type B
    Automatic calibration: at 5 temperature points
    External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
    Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric powert: 300 VA
    Dimensions: 160 x 360 x 350 mm
    Weight incl. standard accessories: 11.0 kg

    All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
    The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 40 mm.

bottom of page