The MI 6601 MediTest can be used as stand-alone tester without a PC or laptop in the field. For office and laboratory use, the tester can work in combination with Metrel Medical Electrical Safety Manager (MMESM). It supports creation and execution of test sequences in accordance with IEC/EN 60601 and asset management. Reports are created with integrated online services Metrel Cloud Reports and Metrel Cloud Storage. Compliance with IEC/EN 60601 is a widely accepted benchmark and a requirement for commercialisation of electrical medical equipment around the world. In Metrel we believe that our new MI 6601 MediTest is the fastest and easiest-to-use electrical safety compliance tester on the market, which entirely covers prescribed standard test procedures, without any compromises or simplifications.
The most complete IEC 60601 tester on the market;
- 10 universal configurable connections. They can be configured as applied parts, non-earthed parts, earthed parts;
- Dedicated test ports for connection to functional earthing and signal I/O connections;
- All possible configurations acc. to the IEC 60601 and IEC 62353 standards;
- Tests can be run on the instrument directly or via a PC;
- Fully automatic test flow with Auto Sequences®;
- Tests and limits are automatically set acc. to the set configuration of medical equipment;
- Test sequences are optimized for fastest work flow;
- Supports measuring by standards IEC/EN 60601 (2nd and 3rd editions), IEC/EN 62353, ANS/NZS 3551, EN 50678 and EN 50669;
- Use of single tests to diagnose problems easily;
- High current for testing continuity of PE connections: up to 25 A;
- AC and DC values of patient leakage and patient auxiliary current;
- Communication over USB, RS232 and Ethernet;
- Input data in any desired way: touchscreen, barcode/QR code scanner, wireless keyboard, PC SW; Optional printers or NFC writers for creating labels and tags;
- New subscription-based software solution including MMESM, Metrel Cloud Reports and Metrel Cloud storage;
- IP 40 case open, IP 65 case closed;
- Carrying case and bag for accessories
- Standard warranty 2 years.
Metrel MI 6601 MediTest Medical Equipment Safety Tester
Order Info
Please allow 3 - 4 weeks lead time for this new product to arrive.
Comes with a 2 Year (24 month) Warranty from the manufacturer.
This product is only available to customers in North America and Canada.
Contact us for any questions or more information about this product.
Features
- PE resistance with 200 mAAC and 25 AAC;
- Insulation resistance with test voltage 250 VDC or 500 VDC.
- All test configurations acc. to IEC 62353 are supported;
- Leakage current measurements with 1 μA resolution; AC, DC and TRMS value of leakage currents;
- All leakage current measurements as defined in IEC 60601-1 standard (patient, auxiliary, earth, touch);
- All leakage current measurements as defined in IEC 62353 (equipment and applied part; alternative, direct, differential method);
- Measurement according to portable appliance standards EN 50678 and EN 50699;
- Point-to-point, touch and mains voltage tests;
- Equipment power;
- IEC lead test;
- Pre-set or configurable visual and functional inspections.
Included Accessories
- Instrument MI 6601 MediTest
- A 1080 Mains cable
- A 1758 Test lead, black, 1 m
- A 1759 Test lead, brown, 1 m
- A 1760 Test lead, green, 1 m
- A 1761 Test lead, yellow, 1 m
- A 1762 Test lead, violet, 1 m
- A 1014 Test probe, black
- A 1298 Test probe, brown
- A 1062 Test probe, green
- A 1013 Crocodile clip, black, 2pcs
- A 1297 Crocodile clip, brown
- A 1309 Crocodile clip, green
- A 1546 Crocodile clip, yellow
- A 1727 USB cable
- A 1017 Communication cable RS232
- A 1500 Bag for accessories
- Subscription to Metrel Medical Software solution: Metrel Medical ES Manager, Metrel Cloud Reports, and Metrel Cloud Storage.